access icon openaccess Study on the performance of X-ray machines used in K-fluorescent devices

Since Roentgen discovered X-rays, X-rays have been widely used in medical diagnosis and industrial non-destructive testing. With the improvement of the performance of aviation technology and instrumentation, higher requirements have been set for the calibration of detectors. X-ray machine performance is very important for calibration. Tube voltage, tube current, and inherent filtration are important parameters of the X-ray machine and directly affect the image imaging quality and radiation dose. Here, the voltage performance of the X-ray machine was measured using HPGe detector. Measure the tube voltage of the X-ray machine using the end of the spectrum method, it is shown that the performance of the X-ray machine shows the tube voltage within ±2%. Measurement of the linearity of the X-ray tube current and current stability using the PTW30013 ionisation chamber. The relative deviation of the actual measured tube current is 0.08%. The inherent filtration of the X-ray machine was measured by the half-value layer method, and the inherent filtration was 0.058 mm Al. The inherent filtration of the K-fluorescent device was measured to be 1.487 mm Al.

Inspec keywords: dosimetry; ionisation chambers; calibration; germanium radiation detectors; diagnostic radiography; X-ray tubes

Other keywords: inherent filtration; X-ray tube current; tube voltage; X-ray machine performance; medical diagnosis; HPGe detector; K-fluorescent devices; PTW30013 ionisation chamber

Subjects: X-rays and particle beams (medical uses); X-ray techniques: radiography and computed tomography (biomedical imaging/measurement); Ionisation chambers; X-ray and gamma-ray equipment; Radiation dosimetry in medical physics; Patient diagnostic methods and instrumentation; Radiation protection and dosimetry; X-ray, gamma-ray instruments and techniques; Particle and radiation detection and measurement

References

    1. 1)
      • 6. Zhou, H.S., Luo, J.H., Ma, G., et al: ‘The foundation of quasi-mono-energy continuous hard x-ray unit’, Nucl. Electron. Detect. Technol., 2015, 35, (9), pp. 6064, doi:10.3969/j.issn.0258–0934.2015.09.014.
    2. 2)
      • 7. Larson, H.V., Myers, I.T., Roesch, W.C.: ‘Wide-beam fluorescence X-ray source’, Nucleonic, 1955, 13, (11), p. 100.
    3. 3)
      • 13. Li, D.H., Huang, C.W., Wu, H.L.: ‘Calibration method of γ-ray reference radiation field with low air kerma rate’, Acta Metrologica Sin., 2011, 32, (6A), pp. 1821.
    4. 4)
      • 4. Chapman, D., Thomlinson, W., Johnston, R.E., et al: ‘Diffraction enhanced x-ray imaging’, Phys. Med. Biol., 1997, 42, (11), pp. 20152025.
    5. 5)
      • 10. Zhou, X.: ‘Development of the ground calibration facility and study of the calibration method for the hard X-ray detectors on board HXMT’ (Chinese Academy of Sciences University, Hengyang, 2015).
    6. 6)
      • 9. Wu, J.J., Guo, B., Li, B.: ‘Research on tube potential of industrial X-rays generator by invasive measurement’, Acta Metrologica Sin., 2015, 36, (5), pp. 550554.
    7. 7)
      • 11. Zhang, S., Wu, J.J., Wu, C., et al: ‘HPGe detector energy-efficiency calibration and simulation’, Nucl. Electron. Detect. Technol., 2016, 36, (5), p. 571.
    8. 8)
      • 3. Boone, J.M., Seibert, J.A.: ‘A figure of merit comparison between bremsstrahlung and monoenergetic x-ray sources for angiography’, J. X-Ray Sci. Technol., 1994, 4, (4), pp. 334345.
    9. 9)
      • 8. Kathren, R.L., Rising, F.L., Larson, H.V.: ‘K-fluorescence X-ray: a multi-use tool for health physics’, Health Phys., 1971, 21, (2), pp. 285293.
    10. 10)
      • 1. Wang, J., Kang, X., Wu, J.J.: ‘The measurement and research of (10–200) KeV monoenergetic X-rays’ (University of South China, Hengyang, 2016).
    11. 11)
      • 2. Boone, J.M., Seibert, J.A.: ‘A comparison of mono and polyenergetic x-ray beam performance for radiographic and fluoroscopic imaging’, Med. Phys., 1994, 21, (12), pp. 18531863.
    12. 12)
      • 15. Li, C.Z., Wu, J.J., Peng, A.G., et al: ‘Study on the property of high energy X-ray facility’, Nucl. Electron. Detect. Technol., 2017, pp. 569574.
    13. 13)
      • 5. Campbell, J.L., Millman, B.M., Maxwell, J.A., et al: ‘Analytic fitting of monoenergetic peaks from Si(Li) X-ray spectrometers’, Nucl. Instrum. Meth. B, 1985, 9, (1), pp. 7179.
    14. 14)
      • 16. International Organization for Standardization, X and Gamma Reference Radiation for Calibrating Dosemeters and Doserate Meters and for Detemining their Response as a Function of Photon Energy –part 1: Radiation Characteristics and production methods, ISO4037-1[S], 1996.
    15. 15)
      • 14. Li, C.Z., Wu, J.J., Peng, A.G., et al: ‘Study on the method of K fluorescence X-ray photon fluence measurement’, Acta Metrologica Sin., 2016, 37, (z1), pp. 10001158.
    16. 16)
      • 12. Hua, Y., Zhu, Z.B., Liu, Y.Q., et al: ‘Calibration of HPGe detector efficiency’, Nucl. Electron. Detect. Technol., 2014, pp. 8688.
http://iet.metastore.ingenta.com/content/journals/10.1049/joe.2018.9104
Loading

Related content

content/journals/10.1049/joe.2018.9104
pub_keyword,iet_inspecKeyword,pub_concept
6
6
Loading