© The Institution of Engineering and Technology
The authors have previously demonstrated how a transmitter (Tx), a reciprocal transmitter/receiver (Tx/Rx) signal path and two unidirectional receiver (Rx) paths can be used together with short, open and load standards for the absolute vector error correction (AVEC) of a Tx/Rx module. Once calibrated, this Tx/Rx module can then provide accurate vector measurements of the signals that are flowing into and/or out of the test port. In order to simplify the analysis, the AVEC technique was applied to a simplified baseband circuit that did not include frequency conversion mixers in a previous paper. Now, in this paper the authors first show how the AVEC technique can be extended to the vector calibration of high-frequency receivers that involve frequency conversion mixers. The authors then show how to calibrate a system that allows for wideband absolute phase relationship measurements of periodic modulated signals, provided that the same local oscillator is employed for the two down-conversion receivers, and different radio frequencies and intermediate frequencies are employed in these receivers. This novel AVEC technique is one of the key concepts in the design of a wideband absolute vector signal measurement system, which overcomes the limitations of traditional measurement instruments by combining the features of vector signal analysers, spectrum analysers and vector network analysers.
References
-
-
1)
-
S.L. Dvorak ,
B.K. Sternberg
.
New absolute vector error correction technique for a transmitter/receiver module.
IET Sci. Meas. Technol.
,
5 ,
359 -
366
-
2)
-
D.F. Williams ,
F. Ndagijimana ,
K.A. Remley ,
J.A. Dunsmore ,
S. Hubert
.
Scattering-parameter models and representations for microwave mixers.
IEEE Trans. Microw. Theory Tech.
,
1 ,
314 -
321
-
3)
-
Blockley, P., Gunyan, D., Scott, J.B.: `Mixer-based, vector-corrected, vector signal/network analyzer offering 300 kHz–20 GHz bandwidth and traceable phase response', 2005 IEEE MTTT-S Microwave Symp. Digest, June 2005, p. 1497–1500.
-
4)
-
Maury Microwave: ‘Large-signal network analyzer’, MT4463, Data sheet.
-
5)
-
J. Verspecht
.
Large signal network analysis.
IEEE Microw. Mag.
,
82 -
92
-
6)
-
Verspecht, J.: `Calibration of a measurement system for high-frequency nonlinear devices’', 1995, Doctoral, Vrije Universiteit Brussel, available at http://www.janverspecht.com.
-
7)
-
Van den Broeck, T., Verspecht, J.: `Calibrated vectorial nonlinear-network analyzers', IEEE MTT-S Int. Microwave Symp. Digest, 1994, WE3F-38.
-
8)
-
Agilent Technologies: ‘PNA microwave network analyzers – mixer transmission measurements using the frequency converter application’, Application Note 1408-1, 2003.
-
9)
-
Agilent: ‘PSA spectrum analyzers’, Data Sheet, 2007.
-
10)
-
B.K. Sternberg ,
S.L. Dvorak
.
New vector signal measuring system, featuring wide bandwidth, large dynamic range, and high accuracy.
IET Sci. Meas. Technol.
,
4 ,
219 -
232
-
11)
-
El Yaagoubi, M., Neveuxm, G., Barataud, D., Nebus, J.M., Verspecht, J.: `Accurate phase measurements of broadband Multitone signals using a specific configuration of a large signal network analyzer', 2006 IEEE MTT-S Int. Microwave Symp. Digest, 2006, p. 1448–1451.
-
12)
-
B.K. Sternberg ,
S.L. Dvorak
.
Quasi-reciprocal mixer as an absolute vector signal standard at microwave frequencies.
IET Sci. Meas. Technol.
,
4 ,
258 -
267
-
13)
-
Myslinski, M., Schreurs, D., Remley, K., Nauwelaers, B.: `Wideband large-signal RF measurements applied to behavioral model extraction', Proc. 36th European Microwave Conf., September 2006, Manchester, UK.
-
14)
-
Agilent Technologies: ‘Optimizing spectrum analyzer amplitude accuracy’, Application Note 1316, 2000.
http://iet.metastore.ingenta.com/content/journals/10.1049/iet-smt_20070118
Related content
content/journals/10.1049/iet-smt_20070118
pub_keyword,iet_inspecKeyword,pub_concept
6
6