access icon free Theoretical and experimental investigation of six-degree-of-freedom force/thrust measurement stand

Thrust testing units with a piezoelectric dynamometer have unique traits such as excessive stiffness, tremendous measurement accuracy, dynamic performance and no hysteresis. These are widely used in the applications requiring force/thrust measurements in the aerospace industry and high-end tool condition monitoring. The performance of these units is necessary to be evaluated. In this study, an improved layout of six-degree of freedom force/thrust measurement stand is proposed and analysed theoretically and experimentally. The measurement stand is a structural component to measure six components of force, such as axial force/thrust (FX , FY , and FZ ) and other components (MX , MY , and MZ ). Test stand consists of seven piezoelectric sensors in two sections. The front part consists of four piezoelectric force sensors, while the rear part consists of three sensors. The rear section is hexagonal, with three sensors mounted at 120°. The measurement stand can measure the principal force/thrust up to 50,000 N. A mathematical model is derived for every sensor against forces in all directions. To calibrate the stand, a calibration platform is designed and fabricated. The calibration platform can generate a range of forces/moments. Calibration experiments verifies that the measurement stand is fairly functional to measure variety of forces/moments with high repeatability.

Inspec keywords: piezoelectric transducers; force measurement; force sensors; mathematical analysis; dynamometers; calibration

Other keywords: high-end tool condition monitoring; calibration; mathematical model; aerospace industry; piezoelectric force sensors; six-degree-of-freedom force-thrust measurement stand; piezoelectric dynamometer

Subjects: Measurement of mechanical variables; Sensing devices and transducers; Piezoelectric devices; Mechanical variables measurement; Measurement standards and calibration; Sensing and detecting devices; Mathematical analysis; Measurement standards and calibration

References

    1. 1)
      • 11. Lawless, I.M., Ding, B., Cazzolato, B.S., et al: ‘Adaptive velocity-based six degrees of freedom load control for real-time unconstrained biomechanical testing’, J. Biomech., 2014, 47, (12), pp. 32413247.
    2. 2)
      • 5. Brimhall, Z.N., Divitotawela, N., Atkinson, J.P., et al: ‘Design and validation of a six degree of freedom rocket motor thrust stand’. 44th AIAA/ASME/SAE/ASEE Joint Propulsion Conf. & Exhibit, Hartford, CT, USA, July 2008, pp. 17.
    3. 3)
      • 7. You, A., Be, M.A.Y., In, I.: ‘Thrust stand for low-thrust liquid pulsed rocket engines’, 2010, 095102.
    4. 4)
      • 12. Miloš, P., Davidović, N., Jojić, B., et al: ‘A novel 6 DOF thrust vector control test stand’, Teh. Vjesn. – Tech. Gaz., 2015, 22, (5), pp. 12471254.
    5. 5)
      • 9. Kelly, B.P., Bennett, C.R.: ‘Design and validation of a novel Cartesian biomechanical testing system with coordinated 6 DOF real-time load control: application to the lumbar spine (L1-S, L4-L5)’, J. Biomech., 2013, 46, (11), pp. 19481954.
    6. 6)
      • 13. Xing, Q., Li, T., Zhang, J., et al: ‘Study of the dynamic performance of a thrust stand for small-thrust liquid-pulsed thrusters study of the dynamic performance of a thrust stand for small-thrust liquid-pulsed thrusters’, 2019, 065113.
    7. 7)
      • 8. Wright, A.M., Wright, A.B., Born, T., et al: ‘A six-degree-of-freedom thrust sensor for a lab-scale hybrid rocket’, Meas. Sci. Technol., 2013, 24, (12), p. 125104.
    8. 8)
      • 6. Sun, C.Y.G.B.Y.: ‘Investigation of the piezoelectric thrust test stand for a rocket engine’, IET Sci. Meas. Technol., 2009, 3, (6), pp. 395402.
    9. 9)
      • 10. Control, V.D., Degrees, S., Device, A.R.A.: ‘〜 力, 剛性, 粘性制御精度の検証 〜 the ankle-foot rehabilitation assistive device’, 2014, 52, pp. 8896.
    10. 10)
      • 2. Kim, G.: ‘Force/moment sensor testing machine and the development of a six-component force/moment sensor testing machine and evaluation of its uncertainty’, 2000.
    11. 11)
      • 3. Kim, G.S.: ‘The design of a six-component force/moment sensor and evaluation of its uncertainty’, Meas. Sci. Technol., 2001, 12, (9), pp. 14451455.
    12. 12)
      • 1. Change, N.D.: ‘Introduction to piezoelectric force sensors’. DYTRAN General Catalogue & Instrumentation Handbook, 1995, pp. 7072.
    13. 13)
      • 4. Brimhall, Z., Atkinson, J., Kirk, D., et al: ‘Design of a novel six degree of freedom solid rocket motor test stand’. 43rd AIAA/ASME/SAE/ASEE Joint Propulsion Conf. & Exhibit, Cincinnati, OH, USA, July 2007, pp. 17.
http://iet.metastore.ingenta.com/content/journals/10.1049/iet-smt.2020.0046