A multilayer piezoelectric actuator has been evaluated for use in a hearing implant. Its frequency response was flat up to 20 kHz. Initial in-vitro trials gave stapes displacement from stimulation at 3.5 V RMS, which is equivalent to that from acoustic stimulation at 90 dB SPL below 1 kHz, increasing to 120 dB at 8 kHz. This performance meets accepted clinical specifications.