The radio frequency coils in magnetic resonance imaging (MRI) systems induce scattered electric fields in the implanted medical lead and result in tissue heating. A novel structure of the medical lead has been designed to reduce the scattered field. Proposed metal nails are placed along the medical lead. The scattered electric field is calculated near the medical lead tip at 64 MHz (1.5 T MRI). By installing this design on the conventional lead, a significant decrease in the scattered electric field at the tip of the lead is observed.
References
-
-
1)
-
3. Mattei, E., Triventi, M., Calcagnini, G., Censi, F., Kainz, W., Mendoza, G., Bassen, H.I., Bartolini, P.: ‘Complexity of MRI induced heating on metallic leads: experimental measurements of 374 configurations’, Biomed. Eng. Online, 7, 2008 (doi: 10.1186/1475-925X-7-11).
-
2)
-
1. Martin, E.T., Coman, J.A., Shellock, F.G., Pulling, C.C., Fair, R., Jenkins, K.: ‘Magnetic resonance imaging and cardiac pacemaker safety at 1.5-tesla’ J. Am. Coll. Cardiol., 2004, 43, pp. 1–10 (doi: 10.1016/j.jacc.2003.12.016).
-
3)
-
4. Yeung, C.J., Susil, R.C., Atalar, E.: ‘RF safety of wires in interventional MRI: using a safety index’, Magn. Reson. Med., 2002, 47, pp. 187–193 (doi: 10.1002/mrm.10037).
-
4)
-
2. Park, S.M., Kamondetdacha, R., Nyenhuis, J.A.: ‘Calculation of MRI-induced heating of an implanted medical lead wire with an electric field transfer function’, J. Magn. Res. Imag., 2007, 26, (5), pp. 1278–1285 (doi: 10.1002/jmri.21159).
-
5)
-
5. Kainz, W.: ‘SAR intercomparison protocol for 1.5 T MR systems draft US’. Food and Drug Administration, Center for Devices and Radiological Health, Office of Science and Engineering Laboratories, 2006.
-
6)
-
S.M. Park ,
R. Kamondetdacha ,
J.A. Nyenhuis
.
Calculation of MRI-induced heating of an implanted medical lead wire with an electric field transfer function.
J. Magn. Res. Imag.
,
5 ,
1278 -
1285
-
7)
-
W. Kainz
.
(2006)
SAR intercomparison protocol for 1.5 T MR systems draft US.
-
8)
-
E. Mattei ,
M. Triventi ,
G. Calcagnini ,
F. Censi ,
W. Kainz ,
G. Mendoza ,
H.I. Bassen ,
P. Bartolini
.
Complexity of MRI induced heating on metallic leads: experimental measurements of 374 configurations.
Biomed. Eng. Online
-
9)
-
C.J. Yeung ,
R.C. Susil ,
E. Atalar
.
RF safety of wires in interventional MRI: using a safety index.
Magn. Reson. Med.
,
187 -
193
-
10)
-
E.T. Martin ,
J.A. Coman ,
F.G. Shellock ,
C.C. Pulling ,
R. Fair ,
K. Jenkins
.
Magnetic resonance imaging and cardiac pacemaker safety at 1.5-tesla.
J. Am. Coll. Cardiol.
,
1 -
10
http://iet.metastore.ingenta.com/content/journals/10.1049/el.2012.4033
Related content
content/journals/10.1049/el.2012.4033
pub_keyword,iet_inspecKeyword,pub_concept
6
6
Correspondence
This article has following corresponding article(s):
in brief