access icon free Bio-signal acquisition circuit with high signal-to-noise ratio for u-healthcare system

The testing of a monolithic bio-signal acquisition circuit with an analogue negative feedback loop for ubiquitous healthcare (u-healthcare) and the monitoring of physical activity is described. The large low-frequency signal is effectively rejected by the proposed circuit, allowing the dynamic range of the more valuable signals to be improved. The proposed circuit has been designed and fabricated using a 0.18 μm 1-poly 6-metal CMOS process. Using an analogue negative feedback loop with a dual-gain filter, the proposed method provides an average signal-to-noise ratio 8.1 dB higher than that of the conventional method.

References

    1. 1)
    2. 2)
      • 1. Gang, G.W., Min, C.H., Kim, T.S.: ‘Design of bio-signal based physical activity monitoring systems’. IEEE EMBS, Int. Conf. Biomedical and Health Informatics, Hong Kong, January 2012, pp. 144147.