The introduction of digital recording system in the high-voltage measuring technique has a great influence on the measuring technique and the evaluation procedures. The formerly used analogue recording instruments were developed for high-voltage measurements and in particular for impulse measurements under very noisy conditions or in other words for high electromagnetic interferences. One of the best measures against the electromagnetic interference was the high signal-to-noise ratio, reached by a high signal voltage level up to 1,500 V for impulse voltage measurements. This high signal level leads also to a high deflection level of about 100 V/cm and requires no amplifier within the analogue oscilloscope.